What Is ICP-MS Testing?

Short answer

ICP-MS (inductively coupled plasma mass spectrometry) is a test for metals and other elements. The sample is dissolved in acid, sprayed into an argon plasma that turns its atoms into ions, and a mass spectrometer sorts and counts those ions by mass. It measures elements such as lead, arsenic, cadmium and mercury at trace levels, which is why drug, food and water standards rely on it.

How ICP-MS works

  1. Digestion. The sample is broken down in acid, often by microwave heating in a closed vessel, so every element ends up in solution (FDA EAM).
  2. Ionization. The solution is sprayed into an argon plasma, which turns the elements into ions.
  3. Mass separation. A mass spectrometer separates the ions by mass-to-charge ratio, so each element is counted on its own.
  4. Quantitation. Results are read against calibration standards, with internal standards to correct for drift and matrix effects (EPA Method 200.8).

What ICP-MS is used for

Drug products. FDA’s guidance implements ICH Q3D and USP <232> and <233>, which set limits for elemental impurities in drug products and describe how to test for them (FDA guidance). Q3D places arsenic, cadmium, mercury and lead in Class 1: human toxicants that often come from mined excipients (ICH Q3D(R2)). Impurities can also come from equipment and container closures.

Food and water. FDA’s Elemental Analysis Manual uses ICP-MS for lead, cadmium, arsenic, mercury and other elements in food (FDA EAM), and EPA Method 200.8 uses it for trace elements in ground, surface and drinking waters (EPA).

ICH Q3D permitted daily exposures for Class 1 elements
ElementOral limit (µg/day)Parenteral limit (µg/day)
Cadmium52
Lead55
Arsenic1515
Mercury303

Source: ICH Q3D(R2) permitted daily exposures (FDA). Limits are per day of drug taken, so the allowed concentration depends on the daily amount.

ICP-MS vs ICP-OES vs HPLC

ICP-OES also uses a plasma but measures the light each element gives off; ICP-MS counts ions by mass and generally reaches lower detection limits. USP <233> describes both: Procedure 1 is ICP-OES and Procedure 2 is ICP-MS (USP <233>). HPLC and LC-MS are different tools. They separate and measure molecules, such as the active drug in a potency test. An HPLC assay can’t tell you how much lead a product has, and ICP-MS can’t tell you its potency.

Who uses ICP-MS testing?

  • Compounding pharmacies and 503B outsourcing facilities checking APIs, excipients or finished products for elemental impurities.
  • Businesses checking raw materials, supplements or cosmetics for heavy metals. FDA notes its food method isn’t validated for mineral supplements or cosmetics without extra verification (FDA EAM), so each sample type needs a checked method.
  • Consumers who want to know whether a product contains heavy metals.

Testing at Renew Lab Group

Our Houston lab runs ICP-MS alongside HPLC and LC-MS potency and purity testing, toxicology, LAL endotoxin, microbiology, TOC and extractables/leachables. See our HPLC testing lab, compounding pharmacy potency testing, testing for individuals, toxicology testing and lab testing FAQ. To request testing, call 936-379-5515 or email info@renewlabgroup.com.

Related questions

What is the difference between ICP-MS and ICP-OES?

Both use an argon plasma. ICP-OES measures emitted light; ICP-MS counts ions by mass and generally detects lower levels. USP <233> lists both as procedures for elemental impurities (USP <233>).

Is ICP-MS a type of chromatography?

No. ICP-MS measures elements, not molecules, and has no separation column. It can be paired with HPLC, called HPLC-ICP-MS, to separate different chemical forms of one element, such as arsenic species.

Can ICP-MS test supplements for heavy metals?

Yes, with a method checked for that sample type. FDA says its food ICP-MS method needs extra verification before use on mineral supplements or cosmetics (FDA EAM).

Can Renew Lab Group run ICP-MS on my product?

Yes. Our Houston lab tests products for compounding pharmacies, businesses and consumers. For a quote, call 936-379-5515 or email info@renewlabgroup.com.

Talk to the lab about ICP-MS testing

Tell us what you need tested and we’ll recommend a method and send a quote. Contact us at 936-379-5515 or info@renewlabgroup.com for current pricing, turnaround and sample requirements. Samples can be dropped off at our lab at 6001 Central Crest, Houston, TX 77092.

Call 936-379-5515Email the labRequest testing

Sources

Leave a Comment

Your email address will not be published. Required fields are marked *

Part of the Rhett Kenagy family of brands: Kynetide  |  Rhett Kenagy